Analytical Suite Metadata

Chemex AAS (Mo,V (dl=10); nitrous oxide acetylene)



Chemex, atomic absorption spectroscopy - nitrous oxide acetylene (molybdenum; vanadium, dl=10ppm)

General Methodology:

Determination was by atomic absorption spectroscopy using a nitrous oxide acetylene flame.  A 0.5 gram sample was reacted with 1.5 ml concentrated HNO3 at 90oC for 30 minutes.  At this point 0.5 ml concentrated HCl was added and the digestion was continued at 90oC for an additional 90 minutes.  After cooling, 8 ml of 1250 ppm Al solution were added and the sample solution was diluted to 10 ml before aspiration.

Instrument: unstated

Sample material: solid

Sample weight: unstated


Analytical Technique:

Hierarchical classification

Technique >> Spectroscopy >> Electromagnetic radiation >> Visible-ultraviolet >> Atomic >> Absorption >> Flame >> C2H2/N2O

Children

None


Analytical Decomposition:

Hierarchical classification

Decomposition >> Aqueous dissolution >> Acid >> Moderate Acid >> Aqua Regia (s.l.)

Children

Aqua Regia (s.s.) | Lefort | Aqua Regia (Acme/BVM) | Aqua regia (NGR) | Aqua regia (NASGLP)


Analytical Methods:

IndexMethod OrderMethod IDQuantityDetection Limit
1  1009     Mo_AAS  2
2  1015     V_AAS  10

Associated Analytical Packages:

IndexPackage IDPackage Name
1  294     Chemex (NGR lake sed 1983 samples) AAS (8+4), CVAAS (Hg), LOI500, U, ISE( F)
Date modified: