Analytical Suite Metadata

Chemex (surface water) alkalinity – d.l. = 2



Chemex, surface water; alkalinity – detection limit is 2 ppm

General Methodology:

Alkalinity in waters was determined by titrating a 25 ml aliquot of the sample with 0.02 N H2SO4 using a Corning combination electrode and a Corning model 135 pH meter.  The end point was pH 4.5.

Instrument:

Sample material: aqueous

Sample weight: unstated


Analytical Technique:

Hierarchical classification

Technique >> Titration

Children

None


Analytical Decomposition:

Hierarchical classification

Decomposition >> None

Children

None


Analytical Methods:

IndexMethod OrderMethod IDQuantityDetection Limit
1  2198     Alk  2

Associated Analytical Packages:

IndexPackage IDPackage Name
1  425     Chemex (NGR surface water) LIF(U), ISE(F), ICP (Ca, Mg), pH and Alk(d.l.=2)