Analytical Suite Metadata

Acme F



Acme F by ISE

General Methodology:

Fluorine was determined using 0.25-gram sample splits weighed into Ni crucibles.  One gram of sodium hydroxide was added and the mixture was fused at 600oC in a muffle furnace.  The fusion product was dissolved in 7 ml of deionised water and 5 ml of 30% sulphuric acid.  The solutions were transferred to plastic beakers and5 ml of 30% ammonium acetate added.  The volume was made up to 90 ml with deionised water.  The pH was tested and adjusted to 7.8 with either sodium hydroxide or sulphuric acid.  Fluorine content was determined using a fluorine selective ion electrode.

(description taken from GSC Open File 1790)

Instrument: unstated

Sample material: solid

Sample weight: unstated


Analytical Technique:

Hierarchical classification

Technique >> Electrochemical techniques >> Potentiometry >> Ion-specific electrodes

Children

Glass electrode


Analytical Decomposition:

Hierarchical classification

Decomposition >> Fusion >> Alkaline >> Hydroxide >> NaOH

Children

None


Analytical Methods:

IndexMethod OrderMethod IDQuantityDetection Limit
1  1418     F_ISE  10

Associated Analytical Packages:

IndexPackage IDPackage Name
1  3     Acme 1F-MS + LOI,Sn,F
2  157     Acme 1F-MS + Sn,F
3  482     Acme (NGR stream sediment) IPC-MS / ICP-ES (35 element), LOI, Sn, F
4  483     Acme (NGR stream sediment) IPC-MS / ICP-ES- HMC (36 element), Sn, F
5  489     Acme (NGR stream sediment) IPC-MS / ICP-ES (35 element), LOI, Sn (LiBO2 fusion), F
6  493     Acme (NGR stream sediment) IPC-MS / ICP-ES (36 elements + Au, Ga, d.l.=0.2; S, d.l.=0.01;W, d.l.=0.2 ppm)