Analytical Suite Metadata

Acme LOI



Acme loss on ignition

General Methodology:

Loss-on-ignition was determined using a one-gram sample.  The sample, weighed into a Leco crucible, was placed in a 100oC muffle furnace and brought up to 500°C for one hour.  The oven was cooled to 100oC and crucibles transferred to a desiccator for cooling to room temperature.  The crucibles were re-weighed, and the difference was reported as loss-on-ignition.

(description taken from GSC Open File 1790)

Instrument: unstated

Sample material: solid

Sample weight: unstated


Analytical Technique:

Hierarchical classification

Technique >> Gravimetry >> Volatilisation >> LOI, 500°C

Children

None


Analytical Decomposition:

Hierarchical classification

Decomposition >> None

Children

None


Analytical Methods:

IndexMethod OrderMethod IDQuantityDetection Limit
1  1416     LOI  1

Associated Analytical Packages:

IndexPackage IDPackage Name
1  3     Acme 1F-MS + LOI,Sn,F
2  482     Acme (NGR stream sediment) IPC-MS / ICP-ES (35 element), LOI, Sn, F
3  489     Acme (NGR stream sediment) IPC-MS / ICP-ES (35 element), LOI, Sn (LiBO2 fusion), F
4  493     Acme (NGR stream sediment) IPC-MS / ICP-ES (36 elements + Au, Ga, d.l.=0.2; S, d.l.=0.01;W, d.l.=0.2 ppm)