Analytical Suite Metadata

Chemex (surface water) Orion ISE (F)



Chemex, surface water; ion-selective electrode (fluorine).  The fluoride was measured using a model 401 Orion specific ion meter with an Orion fluoride electrode.

General Methodology:

Fluoride in water samples was determined using an Orion fluoride electrode and a model 401 Orion specific ion meter.  Prior to measurement, an aliquot of the sample was mixed with an equal volume of a modified TISAB solution (total ionic strength adjustment buffer).  The modification consisted of adding 60 ml 8 M KOH solution to the buffer.  This permitted the re-analysis of fluoride in acidified water samples when required.  When this analysis was required, acidified standard solutions were used for calibration.

Instrument:

Sample material: aqueous

Sample weight: unstated


Analytical Technique:

Hierarchical classification

Technique >> Electrochemical techniques >> Potentiometry >> Ion-specific electrodes

Children

Glass electrode


Analytical Decomposition:

Hierarchical classification

Decomposition >> None

Children

None


Analytical Methods:

IndexMethod OrderMethod IDQuantityDetection Limit
1  1074     F_ISE  20

Associated Analytical Packages:

IndexPackage IDPackage Name
Date modified: