Bondar-Clegg LOI



Loss on ignition (LOI) was determined using a 500 mg sample.  The measurements were done by Bondar-Clegg and Company Ltd. in Vancouver, British Columbia.

General Methodology:

Loss on ignition was determined using a 500 mg sample.  The sample, weighed into a 30 ml beaker, was placed in a cold muffle furnace and brought up to 500°C over a period of 2 - 3 hours.  The sample was left at this temperature for 4 hours, then allowed to cool to room temperature for weighing.

Instrument: unstated

Sample material: solid

Sample weight: unstated


Analytical Technique:

Hierarchical classification

Technique >> Gravimetry >> Volatilisation >> LOI

Children

None


Analytical Decomposition:

Hierarchical classification

Decomposition >> None

Children

None


Analytical Methods:

IndexMethod OrderMethod IDQuantityDetection Limit
1  3445     LOI  1
2  3446     LOIWT1 
3  3447     LOIWT2 

Associated Analytical Packages:

IndexPackage IDPackage Name
1  49     Bondar Clegg INAA + AAS + HY-AAS + F ISE + Hg-CVAAS + U NADNC + LOI